The meter LB-810 is designed for constant water or sewage intensity flow measurement in the open channel with an overflow or with a Venturi tube. The device takes measurements of the damming up of water height using the echo sounder rule (measurement of the echo ultrasound return time). The microprocessor used in the device counts the obtained distance measurement result (damming up of water H) to the amount of flow (Q) basing on the set characteristics of the overflow or tube (Q=f(H)). One can introduce to the device's nonvolatile memory (using the cooperating with the flow meter programmer LB-811) any characteristic of the typical measurement flumes with the Venturi tube (KPV) or with the triangular overfall (KIII). In the non standard cases the user can introduce a different characteristic - defined individually (maximum 19 measurement points). It should be stressed that introducing the measurement table flume to the device's memory is not final and can be performed many times. After programming the flow meter LB-810 the programmer LB-811 is disconnected, what secures the flow meter from any mistaken (or done on purpose) programming it by the user.
The functions realized by the flow meter LB-810:
![]() |
An example measurement station equipped with an intensity flow meter LB-810 |
![]() |
Produced by: LAB-EL Laboratory Electronics 68b, Bodycha Str., 02-495 Warsaw tel: (022) 667 71 18, 0601 209656, fax: (022) 867 53 32 e mail: info@label.com.pl home page: http://www.label.com.pl internet meteorological station: http://www.meteo.waw.pl meteorological station on a mobile phone: wap.meteo.waw.pl Last updated: 09.05.2004 |
![]() |