The panel LB-706 cooperates with various measurement probes (sensors) creating a set for measurement of the following values:
Processed measurement data is presented on a built in LCD screen as well as stored in the internal memory of history measurements, it can also be sent to an external device using a communication interface RS-232. The interface can be connected to a thermal printer, modem LB-430 for radio transmission GSM/ GPRS or a PC computer. The standard equipment includes: a PC program allowing on visualization of the actual reading on the screen, programming device's settings and reading the registered measurements as well as saving them to a hard disc file.
The panel LB-706 is equipped with a clock used for measurement
registry. It also determines when the device is switched on and off in
the energy saving mode. It allows on using the device as a portable,
one station recorder of: temperature (thermograph), humidity
(hygrograph) and optionally air pressure (barograph).
The panel's interface RS-232 allows on entering calibration data during calibration with adjusting. The panel and cooperating with it measurement probes do not have any mechanical regulation elements (i.e. potentiometers, trimmers). Adjusting (calibration, regulation) means sending through a digital interface to device's non volatile memory the calibration data obtained in model climate conditions. It guarantees comfortable adjusting and high calibration accuracy as well as stability of device's meteorological parameters in time.
The user using the accessories program defines the way in which the measurement results are registered in the built in a panel LB-706 memory. The registry configuration is done through:
It is possible to for the panel LB-706 to have installed various types of registry memory. The maximum number of records that can be stored depends on the type of memory and data being registered i.e. in a standard memory version (memory 32 pages) the device can register up to 2600 measurement results: humidity and temperature in an extended memory version (memory 128 pages) more than 10000 measurement results.
The panel is a device designed for processing and visualization of the measurement results. Small size along with being battery powered allow on using it in the field conditions. In stationary conditions the device can be powered from a small power voltage feeder.
The keyboards of the LB-706 and LB-706B panels allow on switching them on and programming parameters as well as browsing registered measurement results.
The panels in versions LB-706G and LB-706GB have simplified
keyboards that allow only on switching them on and off. The panels
LB-706G and LB-706GB are the newer versions of the panels LB-702 and
LB-702B (they replace them fully). Switching from the panel LB-702 to
LB-706G or to LB-706 and from the panel LB-702B to LB-706GB or to
LB-706B does not require changing a hytherograph LB-701.
Hytherorgaph LB-701 cooperating with a panel LB-706 (or with its versions: LB-706G, LB-706B, LB-706GB) is made in a form of a measurement probe, has a cylinder shape and is connected with it using a cable 1...20 m in length. The measurement probe LB-701 includes: a thin layer temperature sensor Pt-1000, thin layer capacitive relative humidity sensor, electronic processing systems and a system of programmed non volatile memory, where all individual probe characteristic are stored as well as the identification and technical data. A replaceable casing, which type depends on what the device is used for, protects the measurement probe sensors. The device in a standard version is equipped with a sensor casing (anti pollution filter) type F1.
The reading panel LB-706 receives signals that had been already processed that include coded information concerning temperature and relative humidity values. It allows on changeable use of different reading panels with the same hytherograph LB-701 without any influence on the measurement results.
The microprocessor from the device's reading panel, basing on the information from the probe (processed measurement signals and read calibration characteristics) computes the following parameters: relative humidity, temperature, dew point temperature and volume water vapor content in the air. Additionally inside the panel version LB-706B there is a pressure sensor and a precise analog- digital converter. All results (temperature, humidity and optionally pressure) are presented on the alphanumeric LCD screen at the same time (2 lines for 16 characters) and can be sent out using a series port.
Hytherograph LB-701 obtained a certificate of type RP T 95 76 issued by the President of the Central Measurement Office in Warsaw as well as a positive expertise from the Central Army Meteorology Office in Zielonka.
Each thermometer - hygrometer LB-701 (upon a request) can have an individual calibration certificate issued by the LAB-EL Company Laboratory. Barometer (upon an additional request) can have a calibration certificate issued by the Air Force Technical Institute in Warsaw.
The methods used when measuring relative humidity and air temperature using a hytherograph are introduced in the description.
The device is also produced in a version for temperature
measurement
(thermometer LB-701T with a panel LB-706 or its versions: LB-706G,
LB-706B, LB-706GB). The thermometer LB-701T has a measurement converter
(in a cylinder case) and an outside temperature sensor Pt1000 (or
Pt100) in a standard or special case, adjusted to the individual user
requirements (i.e. in a form of a rod from the acid resisting steel,
with a special cable). The special version of sensor
casing allows on
measurement of different values in an extended scope of temperature
-200...+550°C). Thermometer LB-701T can be also purchased separately,
as additional equipment for the already had set of LB-701/LB-706. The
thermometer LB-701T cable length amounts to: between an outside
temperature sensor and a converter - 1..40 m, between a converter and a
reading panel - 1..15 m (cable length is defined by the customer).
TEMPERATURE
MEASUREMENT |
||
---|---|---|
Device
version |
Scope
of measurement |
Measurement inaccuracy |
LB-701 |
-40.. +85°C | +/- 0,1°C, +/- 1 last digit |
LB-701T |
-40.. +85°C -150..240°C -200.. +550°C |
+/- 0,1°C, +/- 1 last digit +/- 0,2°C, +/- 1 last digit +/- 1°C, +/- 1 last digit |
HUMIDITY
MEASUREMENT |
||
---|---|---|
Measurement inaccuracy |
+/- 2,0% +/- 1 last digit |
in the scope of 10.. 90% |
+/- 4,0% +/- 1 last digit |
outside of the scope 10.. 90% |
|
Scope of measurement LB-701 |
10.. 95% 10.. 60% 10.. 40% |
for temperatures up to +40°C for temperatures up to +70°C for temperatures up to +80°C |
Scope of measurement LB-701H | 0.. 100% 0.. 70% 0.. 40% |
for temperatures up to +60°C for temperatures up to +70°C for temperatures up to +80°C |
PRESSURE MEASUREMENT (panel LB-706B) | ||
---|---|---|
Measurement inaccuracy | +/- 1,0 hPa | in the scope of temperatures
0..+50°C |
Scope of measurement | 700.. 1100 hPa | in the scope of temperatures
0..+50°C |
MEASUREMENT READINGS | ||
---|---|---|
Parameter |
Scope of readings |
Resolution |
Air temperature |
-40,0.. +85,0°C | 0,1°C or 0,01°C |
Relative humidity | 0,0..99,9% |
0,1% |
Dew point temperature |
-40,0..+100,0°C | 0.1°C |
Volume water vapor content in the air | 0..99999ppm | 1ppm |
Time | months, days, hrs., min | 1 min |
Pressure (LB-706B) | 700...1100 hPa |
0,1 hPa |
POWER SUPPLY | |
---|---|
Battery |
4 batteries R6 (AA) |
Network | stabilized feeder 12V DC |
ENERGY CONSUMPTION (panel + LB-701) | |
---|---|
interface RS232C not active | app. 5,5 mA |
interface RS232C active | app. 13 mA |
SCOPE OF WORK TEMPERATURES | ||
---|---|---|
Device |
Scope |
Comments |
Probe LB-701 |
-40,0.. +85,0°C |
|
Panel LB-706/ LB-706B |
0..+50°C |
RECOMMENDED CONSTANT WORK CONDITIONS *) | |
---|---|
Scope of temperatures |
10..40°C |
Scope of humidity | 20..80% |
Level of environment corrosion
aggressiveness accord. PN-71/H-04651 |
B |
Degree of casing protection |
IP 60 (limited additionally by
the class of filter protection) |
The following versions are available upon a special order:
![]() |
Produced by: LAB-EL Laboratory Electronics 68b, Bodycha Str., 02-495 Warsaw tel: (022) 667 71 18, 0601 209656, fax: (022) 867 53 32 e mail: info@label.com.pl home page: http://www.label.com.pl internet meteorological station: http://www.meteo.waw.pl meteorological station on a mobile phone: wap.meteo.waw.pl Last updated: 11.09.2009 |
![]() |