Call consultant
+48 22 753 60 32
Home » Devices » Thermohygrometers » Hytherograph LB-718

LB-718 Hytherograph with a Resolution of 0.01°C and Modbus RTU Support

LB-718 Thermohygrometer with a Resolution of 0.01°C and Modbus RTU Support

The LB-718 hytherograph is designed for precise measurement of temperature and relative air humidity in rooms. It can work in a dusty environment because the sensors are surrounded by a filter.

The device talks with the measuring system using the industrial RS-485 Modbus RTU communication standard.

Measurement:
  • Air temperature, resolution 0.01°C, uncertainty 0.1°C
  • Relative air humidity, resolution 0.1% RH, uncertainty 2% RH.
Communication:
  • RS-485 Modbus RTU
Power:
  • 7..28V DC 4mA
Operating range:
  • -40.0…+85.0°C, recomended: 0..40°C
  • 10…90% RH, recomended: 20…80% RH
Closure:
  • Polymas, ABS, IP65, 170x64x35mm
LB-718 Thermohygrometer with a Resolution of 0.01°C and Modbus RTU Support
  • measurement: temperature, relative air humidity
  • robust and dust-proof housing
  • Wide operating temperature range
  • RS-485 interface, MODBUS RTU
  • temperature resolution of 0.01°C
  • PC software - LBX SCADA

Description

The LB-718T digital hytherograph is designed for air temperature and relative air humidity measurement in rooms. It may operate in extremely dusty spaces, thanks to the use of air-tight housing and suitable filter protecting the measuring sensor. It is a small, stationary thermometer, powered by an external source. Measurement data are displayed on a two-line LCD display; they may be displayed with a resolution of 0.1% RH and 0.01°C.

To make the measurements, the hytherograph uses data from the external Pt-1000 temperature sensor and the relative humidity capacitive sensor. On the basis of the known non-linear characteristics of the sensors and digital calibration data stored during calibration on the device storage memory, the built-in microprocessor calculates the current measurement results. The results are available in a digital form for the supervising data collection system via the RS-485 interface and MODBUS RTU protocol.

The LB-718 hytherograph is equipped with a digital RS-485 interface that allows the operation of up to 32 devices on one rail with a total length of up to 1,200 m. Thethermometer integrates the widely-known MODBUS RTU protocol, thanks to which the thermometer may operate in many systems based on this standard. In particular, the LB-718 hytherograph may be connected via the DE-311 Ethernet converter with the LBX software, serving as a central point of the LAB-EL monitoring system.

The hytherograph is equipped with jumpers used for setting the MODBUS address in the range of 1…32.

The output of the LB-718 hytherograph is made with screw clamps (4 camps — power, A and B lines), located inside the housing, to which a cable is led through an opening with a diameter of 7 mm.

The hytherograph does not feature mechanical calibration elements (e.g. potentiometers). The hytherograph calibration includes sending calibration data via a digital interface to the instrument’s permanent storage, obtained from the calibrated climatic conditions. It ensures convenience and high calibration accuracy as well as the stability of the device's metrological parameters over time.

In the standard version, the instrument is equipped with a sensor screen (impurities filter), type F3.

Calibration certificates

Each hytherograph, as per order, may feature an individual calibration certificate provided by LAB-EL Calibration Laboratory.

Dedicated software

This hytherograph may additionally come with an advanced software for Windows — the LBX software — enabling:

  • presentation of current readings from devices;
  • alarm monitoring and signalling;
  • presentation of archived data — charts, reports, histograms;
  • data archive management;
  • access for remote clients.

Technical Data


Dane techniczne termohigrometru

TEMPERATURE MEASUREMENT
Measurement uncertainty +/- 0,1 °C  ±1 final digit
Measurement range -40..+85 °C  
HUMIDITY MEASUREMENT
Measurement uncertainty +/- 2,0 % within the range of 10…90% ±1 final digit
+/- 4,0 % outside the range of 10…90%
Measurement range of LB-718 10..95 % for temperatures up to +40°C
10..60 % for temperatures up to +70°C
10..40 % for temperatures up to +85°C
Measurement range of LB-718H 0..100 % for temperatures up to +60°C
0..70 % for temperatures up to +70°C
0..40 % for temperatures up to +85°C
MEASUREMENT READING
Parameter Reading range Resolution
Air temperature
-40,0..+85,0 °C 0.01°C, optionally 0.1°C (configured with a jumper)
Relative humidity 0,0..99,9 % 0,1 %
RECOMMENDED CONTINUOUS OPERATING CONDITIONS *)
Temperature range 0…40°C
Humidity range 20…80%

*) Exceeding the recommended continuous operating conditions (e.g. when installing the device outdoors) requires the use of additional protective measures for the electronic part of the device against the eventuality of water entering the device (protection of the electronics with silicone or usage of additional external housing).

POWER SUPPLY, MECHANICAL DATA
Voltage range 7…28 V DC  
Power consumption approx. 4 mA without a terminator
External dimensions of the housing  170 x 64 x 35 mm   
Material plastics (polymas, ABS)  
Housing protection class IP 65 additionally limited by the protection class of the filter

DIGITAL INTERFACE

RS-485 transmission parameters: 19,200 bps, 8 information bits, parity control — Even, 1 stop bit (19200/8E1) or (configured with a jumper starting with the firmware v1.0.1) 9,600 bps, 8 information bits, no parity control — none, 1 stop bit (9600/8N1)

MODBUS RTU protocol. Interface connection in the form of 4 screw clamps, accessible after removing the device’s cover. Address set with jumpers in the range of 1…32.

Description of available MODBUS registers

INPUT REGISTER (READ SINGLE/MULTIPLE REGISTER)
The value defined as the register address is the same as the one that should appear in the Modbus data package in the address field. Sometimes the PC software adopts the Modicon convention and requires that the input register address be expanded by 30001 more than the value given in this document — e.g. the DEVID register would then have 30001.
Address Designations TYPE Description
0 DEVID WORD Device type identification 0x0718 (hex format)
1 CPTB WORD Firmware version from the viewpoint of operation via MODBUS. In the event that the user’s software recognises firmware of a newer version (FVER register) than the supported one, it should — while reading this register (CPTB) — confirm that it is able to support the device. The version specified in the CPTB register is possibly the oldest version that the current firmware is backward compatible. Description correct for CPTB = 10000.
2 SNUM WORD Serial number of the device
3 FVER WORD Firmware version in the form of X.Y.Z, where: X — tens of thousands of FVER, Y — hundreds of FVER, Z — units of FVER. E.g.: for FVER = 21203, it is 2.12.3
4 CONFIG WORD The value of a variable set by the manufacturer; meaning of bits:
0: humidity measurement present
1: thermometer featuring wide range of temperature (E version: -200…+550)
2: temperature measurement averaging enabled (for the FLOAT type variable)
3: humidity measurement averaging enabled (for the FLOAT type variable)
4: temperature resolution of 0.01°C
5 STATUS WORD State of the device; meaning of bits:
0: ERROR_T — temperature measurement error
1: ERROR_RH — humidity measurement error
2: CFG_ERR — calibration error
Remaining bits feature no meaning
6.7 T FLOAT Temperature
8.9 RH FLOAT Humidity
10 T_WORD WORD Temperature as a number in U2 code with a resolution of 0.1°C (the read-out value should be divided by 10)
11 T_WORD_AVG WORD Temperature — the same as above, but averaging
12 RH_WORD WORD Humidity as a number in U2 code with the resolution of 0.1% (the read-out value should be divided by 10)
13 RH_WORD_AVG WORD Humidity — the same as above, but averaging
14 CAL_YM WORD Date of the last calibration, LSB: year — 2000, MSB: month
15 CAL_DH WORD Date of the last calibration, LSB: day, MSB: hour
16 CAL_M WORD Date of the last calibration, MSB = 0, LSB: minute
DEVICE IDENTIFICATION
Read Device Identification 43/14 BASIC ID

EXTENSIONS

By special order, the following set-ups are available:

  • The LB-718Fn digital thermometer — version of sensors’ cover (according to description), the version is given if it is not F3.

INTERACTION WITH THE DE-311 ETHERNET CONVERTER

LB-718 devices may be connected to the DE-311 converter, thanks to which the measurement data may be monitored by the LBX software via Ethernet.
Connection of the LB-718 with the DE-311 should be made according to the following table:

DE-311 LB-718 DC 7…28 V power unit Signal description
3 A   Data + (B)
4 B   Data - (A)
5 - - Power - / GND
  + + Power +

Microswitch of the DE-311 converter should be in the RS-485 position:

SW1 SW2 SW3
OFF ON ON

Other versions of the LB-718:

See also
See also
Thermometer LB-718TThermometer LB-718T
Digital Thermometer with High Resolution and MODBUS RTU Support
Thermohygrometer LB-523Thermohygrometer LB-523
Wireless relative humidity and temperature recorder, equipped with WiFi and USB interfaces. It is powered from 2 exchangeable rechargeable batteries.
Hytherograph LB-524Hytherograph LB-524
Battery-powered recorder of air temperature, relative humidity, air pressure and light intensity with GSM module for communication with LBX server.